| MaXterileTM Digital Fuzzy-control Autoclaves | |||||
| DAIHAN MaXterileTM Digital Fuzzy-control Autoclaves, Standard & Recorder-type, 47-/60-/80-/100-Lit. | |||||
| Steam Strtilizer, Electronic Door Lock Sys., Lever-type Sliding Door, Steam Condensing Mechanism, Max.2 kgf/cm2, up to 132℃ | |||||
| ■ DAIHAN-brand New Generation Steam Sterilizers | |||||
| ■ with Certificate & Traceability : | |||||
| Controlled by Serial Number, Certificate, Delivery-information, and Traceable Data base System | |||||
| ■ Controlled by Serial Number, Certificate, Delivery-information,and Traceable Data base System | |||||
| ■ Patented Special Steam Condensing Mechanism | |||||
| ■ Built in DAIHAN-brand New Generation Controller : Patented Jog-Dial Control System | |||||
| ■ Electronic Door Lock System for User Safety. | |||||
| ■ Patented Safety and Easy to Use Lever-type Sliding Door | |||||
| ■ Built in Recorder model (only MaXterile 47R/60R/80R/100R models) : Sterilization Cycle Data for Validation and Analyze Sterilization (Temperature, Pressure, Time, Error Status) | |||||
| ■ “CE-MDD (CE 0470 - Medical Device Directive) Certified” (MaXterile 100, MaXterile 100R excluded) | |||||
| ■ PL (Product Liability) Insurance | |||||
| � Ideal for Biotechnology, Clinical, Environmental, Medical and Food Industry. | |||||
| � Innovative Easy-to-Use Digital Fuzzy Controller. | |||||
| � Included 2×Wire Baskets | |||||
| � Top-loading type Sterilizer for Easy Loading of Large Working Volume. | |||||
| � Casters equipped for Excellent Mobility. | |||||
| � Digital LCD with Back-Light Function. | |||||
| � Digital Timer Function ‘- Delay Time’and‘Operation Time’can be Controlled. | |||||
| � Solid / Liquid Mode can be chosen by a customer considering a Sample Phase. | |||||
| � Complete Safety Pressure Mechanism. | |||||
| � Heat-Protective Lid Cover for User Safety. | |||||
| � Water Low Level Detection | |||||
| � Automatic Over Pressure Release Valve ensures User Safety. | |||||
| � Sterilization Error Indication : Error Alarm and Massage in case of Power Failure and Temperature Error | |||||
| (1) Standard-model | MaXterile 47 | MaXterile 60 | MaXterile 80 | MaXterile 100 | |
| Standard-type | |||||
| (2) Recorder-model | MaXterile 47R | MaXterile 60R | MaXterile 80R | MaXterile 100R | |
| Recorder-type : Built in Recording Printer, Sterilization Cycle Data | |||||
| (Date, Temperature, Pressure, Time, Error Status, etc.) | |||||
| Capacity | 47 Lit. | 60 Lit. | 80 Lit. | 100 Lit. | |
| Temp. Range & Accuracy | Ambient +5℃ to 132℃, ±0.5℃ at 121℃ | ||||
| Pressure & Gauge | 1.2 kgf/cm2 at 121℃/ 2.0 kgf/cm2 at132℃, 0 ~ 3 kgf/cm2 | ||||
| Sterilization mode | Solid- or Liquid-mode | ||||
| Heating Power/ Consumption | 2 kW | 3 kW | 2 kW×2 | 2 kW×2 | |
| Timer & Alarm | 60 min, Error status & Timer-end | ||||
| Display | Digital LCD with Back - Light Function | ||||
| Controller | Digital Fuzzy-Control System (Fully Automatic) | ||||
| Door & Closing System | Lever-type Sliding Door / Safety Electronic Door Closing System | ||||
| Safety Circuit | Over-Temp./-Pressure/-Current protection, Sensor Error Detection | ||||
| Printer | Thermal type printer, Record Paper Roll: w58 ×Φ40 mm | ||||
| Material | Stainless steel (#304)-Pressure Vessel, Powder Coated steel-External | ||||
| Basket | 2ea of Stainless steel Wire Baskets are Included | ||||
| Basket Dimension (mm) | Φ270×h240 | Φ320×h240 | Φ370×h240 | Φ420×h240 | |
| Dimension | Inner (mm) | Φ300×h670 | Φ350×h650 | Φ400×h650 | Φ450×h630 |
| (w×d×h) | Outer (mm) | 680×470×880 | 680×470×900 | 780×580×900 | 780×580×900 |
| Packing Size (mm) | 840×610×h1,252 mm | 950×730×h1,352 mm | |||
| Gross Weight (kg) | 117 kg | 120 kg | 170 kg | 170 kg | |
| Power Supply | 1Phase - AC 120V, 60Hz or | 1 Phase - AC 120V, 60Hz or - AC 230V, 50/60Hz | |||
| - AC 230V, 50/60Hz | 3 Phase - AC 380V, 50/60Hz | ||||