Dumatherm N Pro Protein Analysis Gerhardt |
|
|
Dumatherm N Duo Protein Analysis Gerhardt |
|
|
Specifications |
|
|
DUMATHERM® N Pro |
|
|
Sample size |
0,5 mg bis 1 g |
|
Sample capacity |
1 - 64 samples |
|
Analysis time |
3 - 5 min, depending on the sample type and sample size |
|
Recovery rate |
> 99,5 % |
|
Detection limits with Helium |
0,003 - 50 mg N absolut |
|
Detection limit with Argon |
0,1 - 50 mg N absolut |
|
Standard deviation |
< 0,5 % ( with standard sample preparation ) |
|
Operation |
via PC* with DUMATHERM®-Manager |
|
Temp. range Combustion furnace |
400 - 1.100 °C |
|
Temp. range Reduction furnace |
400 - 1.100 °C |
|
Temp. range Desorption furnace |
50 - 350 °C |
|
Required gases and quality |
Helium (alternatively argon), grade 5.0 (99.999%) Oxygen, grade 5.0 (99.999%) Compressed air or nitrogen, grade 2.6 (99.6%, oil- and water-free) |
|
Gas installation |
Pressure regulators for high purity gas; 1/4“ cylindric thread (female) |
|
Inlet pressure helium |
3 - 6 bar |
|
Inlet pressure oxygen |
3 - 6 bar |
|
Inlet pressure compr. air / nitrogen |
3 - 6 bar |
|
Ambient temp. |
15 °C < t < 35 °C |
|
Nominal voltage |
230 VAC |
|
Dimensions W / D / H |
870 x 710 x 860 mm ( closed ) 1108 x 710 x 1118 |
|
( doors and protective hood opened ) |
|
|
Weight ca. |
85 kg |
|
Nominal wattage DUMATHERM® |
1380 W |
|
Nominal wattage autosampler |
90 W |
|
Current max. |
6:00 AM |
|
Wanted digital balance |
Optional, reading precision min. 0,1 mg |
|
Order Infomation |
|
|
14-0400 DT N Pro DUMATHERM® N. 64-place, complete with |
|
|
|
Autosampler for the automatic feeding of up to 64 samples* |
|
Starter Kit for 1.000 combustions |
|
|
Sample preparation tools |
|
|
DUMATHERM ® -Manager, control software* |
|
|
* Computer |
PC / Laptop and printer are not included in the shipment |
|
* Requirements for PC |
Windows 7®64bit, 2 x USB-interface |
|
Specifications |
|
|
DUMATHERM® N Duo |
|
|
Sample size |
0,5 mg bis 1 g |
|
Sample capacity |
max. 2 samples |
|
Analysis time |
3 - 5 min, depending on the sample type and sample size |
|
Recovery rate |
> 99,5 % |
|
Detection limits with Helium |
0,003 - 50 mg N absolut |
|
Detection limit with Argon |
0,1 - 50 mg N absolut |
|
Standard deviation |
< 0,5 % ( with standard sample preparation ) |
|
Operation |
via PC* with DUMATHERM®-Manager |
|
Temp. range Combustion furnace |
400 - 1.100 °C |
|
Temp. range Reduction furnace |
400 - 1.100 °C |
|
Temp. range Desorption furnace |
50 - 350 °C |
|
Required gases and quality |
Helium (alternative Argon), quality grade 5.0 (99,999 %) Oxygen, quality grade 5,0 (99.999 %) Compressed air or nitrogen, quality grade 2.6 (99.6 %, oil- and water free) |
|
Gas installation |
Pressure regulators for high purity gas; 1/4“ cylindric thread (female) |
|
Inlet pressure helium |
3 - 6 bar |
|
Inlet pressure oxygen |
3 - 6 bar |
|
Inlet pressure compr. air / nitrogen |
3 - 6 bar |
|
Ambient temp. |
15 °C < t < 35 °C |
|
Nominal voltage |
230 VAC |
|
Dimensions W / D / H |
870 x 450 x 710 mm ( with autosampler ) |
|
Weight ca. |
65 kg |
|
Nominal wattage DUMATHERM® |
1380 W |
|
Current max. |
6:00 AM |
|
Wanted digital balance |
Optional, reading precision min. 0,1 mg |
|
Order Information |
|
|
14-0003 DT N DUO DUMATHERM® N Duo, 2-place, complete with : |
|
|
|
Autosampler for the manual feeding of max. 2 samples |
|
Starter Kit for 1.000 combustion |
|
|
Sample preparation tools |
|
|
DUMATHERM®-Manager, control software* |
|
|
* Computer |
PC / Laptop and printer are not included in the shipment |
|
* Requirements for PC |
Windows 7®64bit, 2 x USB-interface |