| Survey Meter Gamma-Scout Hand-held Radiation Detector | |||
| The Gamma-Scout hand-held radiation detector or geiger counter is a general purpose survey meter for measuring and detecting ionizing radiation. It can detect alpha, beta, gamma and x-ray radiation.You can use the Gamma-Scout radiation detector to : | |||
| Determine the source of radiation leakages | |||
| Monitor ambient radiation anomalies | |||
| Monitor personal cumulative dose exposure | |||
| Detect radioactive elements in metal scrap | |||
| Check rocks from the earth for radioactivity | |||
| Technical Data, Calibration | |||
| Display | Liquid-crystal display ( LCD ), 4-digit, numeric with description,quasi-analogue logarithmic bar chart | ||
| Operating mode indicators | |||
| Radiation Detector | End-window counting tube to the Geiger-Müller principle Stainless steel housing | ||
| Measuring length 38.1 mm, measuring diameter 9.1 mm | |||
| Mica window 1.5 to 2 mg/cm² | |||
| Zero rate <10 pulses per minute with screening by 3 mm AI and 50 mm Pb operating temperature -20 to +60°C, operating voltage approx. 450 V calibrated scale 0.01 μSv/h to 5,000.00 μSv/h ( above and below this, we refer to as the “display area” ) | |||
| Radiation Types | α | from 4 MeV | |
| ß | from 0.2 MeV | ||
| γ | from 30 keV | ||
| Selection shield | α + ß + γ | without shielding | |
| ( for point radiation ) | ß + γ | AI foil approx. 0.1 mm, shields completely | |
| γ | Al shield approx. 3 mm, shields α completely and ß to 2 MeV, weakens γ by less than 7% based on Cs-137 | ||
| Recycling | Devices can be returned free-of-charge and we will pass them on for recycling | ||
| Power Consumption | Less than 10 microamperes under environmental radiation | ||
| Memory | 256,000 byte ( 100,000 data sets ) | ||
| Housing | Impact-resistant plastic | ||
| Dimensions | Length 163 mm x width 72 mm x height 30 mm | ||
| Interference Protection | European CE standard, US FCC-15 standard | ||
| 
 | Data-reset is not a malfunction | ||
| Technical Data, Calibration | |||
| Physical Parameter | Si unit | Old unit | Relation | 
| Activity | Becquerel (Bq) | Curie (Ci) | 1 Ci = 3.7 * 10¹⁰ Bq | 
| 1 Bq = 1/s | 1 Bq = 2.7 * 10⁻¹¹ Ci | ||
| = 27 pCi | |||
| Ion dose I | Coulomb / kg | Röntgen (R) | 1 R = 2.58 * 10 C/kg | 
| 1 C/kg = 3876 R | |||
| Energy dose D | Gray (Gy) | Rad (rd) | 1 rd = 0.01 Gy | 
| 1 Gy = 100 rd | |||
| Equivalent dose H | sievert (Sv) | Rem (rem) | 1 rem = 0.01 Sv | 
| 1 Sv = 100 rem | |||
| Effective Dose H | sievert (Sv) | Calculated value /radiation protection | |