Sensonic 4000 Gas Analyzers |
|||
General Parematers - Version A ( in soft casing ) |
|||
Dimensions ( W x H x D ) |
460 x 260 x 240 mm |
||
Weight ( without accessories ) |
6.2 ÷ 7.2 kg |
||
Casing material |
Textile ( polyester ) |
||
Operating conditions |
T: 10°C ÷ 50°C, RH: 5% ÷ 90% ( non-condensing ) |
||
Storing temperature |
0°C ÷ +55°C |
||
Power supply: input | max power consumption |
90 ÷ 240 VAC | 70 W |
||
Battery: type | work time | charging time |
Lead-acid, rechargeable 12V / 2.2Ah | 7h | 14h |
||
Data memory: size | number of results |
32kB | 30 reports + 10 banks ( 1024 sets of data ) |
||
Display |
Graphical LCD 128 x 128, with variable contrast and backlighting |
||
Printer |
High-speed dot matrix, graphic printer for 57mm normal paper |
||
Analogue outputs |
Two current ( 0/4mA ÷ 20 mA ) |
||
Gas pump | gas flow |
Diaphragm, max 2 l/min ( with automatic flow control ) | 90l/h ( 1.5l/min ) |
||
Purging pump for CO sensor (optional) |
Diaphragm, max 1.5 l/min |
||
Communication interface with PC computer |
RS-232C |
||
Gas filtering |
1. In-line filter included in the gas probe hose |
||
2. Built-in input filter with water-trap and replaceable insert |
|||
General Parematers - Version B ( in hard casing ) |
|||
Dimensions ( W x H x D ) |
455 x 270 x 220 mm |
||
Weight ( without accessories ) |
8.2 ÷ 9.2 kg |
||
Casing material |
Wood & aluminium |
||
Operating conditions |
T : 10°C ÷ 50°C, RH: 5% ÷ 90% ( non-condensing ) |
||
Storing temperature |
0°C ÷ +55°C |
||
Power supply: input | max power consumption |
90 ÷ 240VAC | 70W |
||
Battery: type | work time | charging time |
Lead-acid, rechargeable 12V / 2.2Ah | 7h | 14h |
||
Data memory : size | number of results |
32 kB | 30 reports + 10 banks ( 1024 sets of data ) |
||
Display |
Graphical LCD 128 x 128, with variable contrast and backlighting |
||
Printer |
High-speed dot matrix, graphic printer for 57mm normal paper |
||
Analogue outputs |
Two current ( 0/4mA ÷ 20mA ) |
||
Gas pump | gas flow |
Diaphragm, max 2l/min ( with automatic flow control) | 90l/h ( 1.5l/min ) |
||
Purging pump for CO sensor (optional) |
Diaphragm, max 1.5l/min |
||
Communication interface with PC computer |
RS-232C |
||
Gas filtering |
1. In-line filter included in the gas probe hose |
||
2. Built-in input filter with water-trap and replaceable insert |
|||
Minidryer ( optional ) |
|||
Dimensions ( W x H x D ) |
24 x 120 x 124 mm |
||
Weight |
800 g |
||
Operating conditions |
T: 10°C ÷ 50°C, RH: 5% ÷ 90% ( non-condensing ) |
||
Storing temperature |
-20°C ÷ +55°C |
||
Power supply : input | max power consumption |
15V DC ( from analyzer's “Probe” socket ) | 10 W |
||
Drying method |
Water condensation by rapid cooling down |
||
Cooler type |
based on Peltier element |
||
Cooling temp. |
Down to +4°C electronically stabilised / Dew point of outlet gas at least 8°C below the ambient air temperature |
||
Maximum gas flow for efficient drying |
90 l/h |
||
Condensate pump |
Peristaltic, 38 ml/min. |
||
Measurements of gas components |
|||
Method |
Range | Resolution |
Accuracy |
Time (T90) |
O2 – Oxygen |
|||
Electrochemical sensor |
20.95% | 0.01% |
± 0.1% abs. or 5% rel |
45 sec |
Electrochemical, partial pressure |
20.95% | 0.01% |
± 0.1% abs. or 5% rel |
45 sec |
Electrochemical, partial pressure |
25.00% | 0.01% |
± 0.1% abs. or 5% rel |
45 sec |
Electrochemical, partial pressure |
100.00% | 0.1% |
± 0.1% abs. or 5% rel |
45 sec |
CO – Carbon monoxide |
|||
Electrochemical sensor |
20 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
Electrochemical with H2 compensation |
2 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
Electrochemical sensor |
10% | 10ppm |
± 50ppm abs. or 5% rel |
45 sec |
NDIR (as a second sensor only) |
10% | 0.01% |
± 0.05% abs. or 5% rel |
45 sec |
NDIR (as a second sensor only) |
100% | 0.1% |
± 0.5% abs. or 5% rel |
45 sec |
CO2 – Carbon dioxide |
|||
NDIR |
25% | 0.01% |
± 0.05% abs. or 5% rel |
45 sec |
NDIR |
50% | 0.01% |
± 0.05% abs. or 5% rel |
45 sec |
NDIR |
100% | 0.1% |
± 0.5% abs. or 5% rel |
45 sec |
CxHy – Total Hydrocarbons |
|||
NDIR |
5% | 0.01% |
± 0.05% abs. or 5% rel |
45 sec |
NDIR |
25%| 0.01% |
± 0.05% abs. or 5% rel |
45 sec |
NDIR |
100% | 0.1% |
± 0.5% abs. or 5% rel |
45 sec |
NO – Nitric oxide |
|||
Electrochemical sensor |
5 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
NO2 – Nitrogen dioxide |
|||
Electrochemical sensor |
1 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
SO2 – Sulphur dioxide |
|||
Electrochemical sensor |
5 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
H2S – Hydrogen sulfide |
|||
Electrochemical sensor |
1 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
Electrochemical sensor |
1 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
H2 – Hydrogen |
|||
Electrochemical sensor |
2 000ppm | 1ppm |
± 10ppm abs. or 5% rel |
45 sec |
Electrochemical sensor |
20 000ppm | 1ppm |
± 10ppm abs. or 5% rel |
45 sec |
NH3 – Ammonia |
|||
Electrochemical sensor |
1 000ppm | 1ppm |
± 10ppm abs. or 5% rel |
45 sec |
CL2 - Chlorine |
|||
Electrochemical sensor |
250ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
HCl - Hydrogen chloride |
|||
Electrochemical sensor |
50ppm | 1ppm |
± 5ppm abs. or 5% rel |
120 sec |
Other measurements / calculated results |
|||
Method |
Range | Resolution |
Accuracy |
Time (T90) |
Tgas – gas temperature |
|||
K-type thermocouple |
-10 ÷ 1000°C | 0.1°C |
± 2°C |
10 sec |
S-type thermocouple |
-10 ÷ 1500°C | 0.1°C |
± 2°C |
10 sec |
Tamb – boiler intake air temperature |
|||
PT500 resistive sensor |
-10 ÷ 100°C | 0.1°C |
± 2°C |
10 sec |
T1 – external temperature |
|||
K-type thermocouple |
-10 ÷ 1000°C | 0.1°C |
± 2°C |
10 sec |
S-type thermocouple |
-10 ÷ 1500°C | 0.1°C |
± 2°C |
10 sec |
T2 – external temperature |
|||
PT500 resistive sensor |
-10 ÷ 100°C | 0.1°C |
± 2°C |
10 sec |
K-type thermocouple |
10 ÷ 100°C | 0.1°C |
± 2°C |
10 sec |
T3 – external temperature |
|||
K-type thermocouple |
-10 ÷ 1000°C | 0.1°C |
± 2°C |
10 sec |
S-type thermocouple |
-10 ÷ 1500°C | 0.1°C |
± 2°C |
10 sec |
T4 – external temperature |
|||
PT500 resistive sensor |
-10 ÷ 100°C | 0.1°C |
0.3m/s abs. or 5% rel |
10 sec |
Differential pressure |
|||
Silicon piezoresistive pressure sensor |
-25hPa ÷ +25hPa | 10Pa (0.01hPa) |
± 2Pa abs. or 5% rel |
10 sec |
Gas flow velocity |
|||
Indirect, with Pitot tube & pressure sensor |
1 ÷ 50m/s | 0.1m/s |
0.3m/s abs. or 5% rel |
10 sec |
Lambda λ – excess air number |
|||
Calculated |
1 ÷ 10 | 0.01 |
± 5% rel |
10 sec |
qA – stack loss |
|||
Calculated |
0 ÷ 100% | 0.1% |
± 5% rel |
10 sec |
Eta η – combustion efficiency |
|||
Calculated |
0 ÷ 120% | 0.1% |
± 5% rel |
10 sec |
U1÷U2 – external analogue input (voltage) |
|||
Delta-sigma ADC |
-20V ÷ +20V | 0.01V |
± 2% rel |
10 sec |
I1÷I2 – external analogue input (current) |
|||
Delta-sigma ADC |
-20mA ÷ +20mA | 0.01mA |
± 2% rel |
10 sec |