Sensonic 4500 Gas Analyzers |
|||
Specifications |
|||
Dimensions ( W x H x D ) |
470 x 310 x 160 mm |
||
Weight ( without accessories ) |
12.0÷12.8 kg |
||
Casing material |
Aluminium padded with foam and fabric ( polyester ) |
||
Operating conditions |
T : 10 °C ÷ 50 °C, RH: 5 % ÷ 90 % ( non-condensing ) |
||
Storing temperature |
0 °C ÷ +55 °C |
||
Power supply : Input | maximal power consumption |
115 VAC or 230 VAC | 90 W (without heated hose) |
||
Battery: type | work time | charging time |
Lead-acid, rechargeable 12V / 2.2Ah | 7h | 14h |
||
Data memory: size | number of results |
32kB | 30 reports + 10 banks (1024 sets of data) |
||
Display |
Graphical LCD 128 x 128, with variable contrast and backlighting |
||
Printer |
High-speed dot matrix, graphic printer for 57mm normal paper |
||
Analogue outputs (optional) |
Two current (0/4÷20mA) or voltage (0÷10V) outputs |
||
Gas pump | gas flow |
Diaphragm, max 2l/min (with automatic flow control) | 90l/h (1.5l/min) |
||
Purging pump for CO sensor |
Diaphragm, max 1.5l/min |
||
Communication interface with PC computer |
RS-232C |
||
Gas filtering |
1. Heated filter included in the heated hose |
||
2. Built-in final filter (behind the gas dryer) with replaceable insert |
|||
Built-in gas dryer, heated hose driver, heated hose |
|||
Drying method |
Water condensation by rapid cooling down |
||
Cooler type |
based on Peltier element |
||
Cooler temperature |
+5°C electronically stabilised |
||
Cooler temperature hysteresis |
~ 1°C |
||
Maximum gas flow for efficient drying |
100l/h |
||
Condensate pump |
Peristaltic, 38ml/min. |
||
Heated hose temperature |
+120°C electronically stabilised |
||
Heated hose temperature hysteresis |
~ 5°C |
||
Heated hose length |
3m (optionally 5m or 10m) |
||
Heated hose power consumption |
360W (max) |
||
Heated hose thermocouple wires |
K-type (S-type optionally) |
||
Measurements of gas components |
|||
Method |
Range | Resolution |
Accuracy |
Time (T90) |
O2 – Oxygen |
|||
Electrochemical sensor |
20.95% | 0.01% |
± 0.1% abs. or 5% rel |
45 sec |
Electrochemical, partial pressure |
20.95% | 0.01% |
± 0.1% abs. or 5% rel |
45 sec |
Electrochemical, partial pressure |
25.00% | 0.01% |
± 0.1% abs. or 5% rel |
45 sec |
Electrochemical, partial pressure |
100.00% | 0.1% |
± 0.1% abs. or 5% rel |
45 sec |
CO – Carbon monoxide |
|||
Electrochemical sensor |
20 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
Electrochemical, with Hâ‚‚ compensation |
2 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
Electrochemical sensor |
10% | 10ppm |
± 50ppm abs. or 5% rel |
45 sec |
NDIR (as a second sensor only) |
10% | 0.01% |
± 0.05% abs. or 5% rel |
45 sec |
NDIR (as a second sensor only) |
100% | 0.1% |
± 0.5% abs. or 5% rel |
45 sec |
CO2 – Carbon dioxide |
|||
NDIR |
25% | 0.01% |
± 0.05% abs. or 5% rel |
45 sec |
NDIR |
50% | 0.01% |
± 0.05% abs. or 5% rel |
45 sec |
NDIR |
100% | 0.1% |
± 0.5% abs. or 5% rel |
45 sec |
CxHy – Total Hydrocarbons |
|||
NDIR |
5% | 0.01% |
± 0.05% abs. or 5% rel |
45 sec |
NDIR |
25%| 0.01% |
± 0.05% abs. or 5% rel |
45 sec |
NDIR |
100% | 0.1% |
± 0.5% abs. or 5% rel |
45 sec |
NO – Nitric oxide |
|||
Electrochemical sensor |
5 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
NO2 – Nitrogen dioxide |
|||
Electrochemical sensor |
1 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
SO2 – Sulphur dioxide |
|||
Electrochemical sensor |
5 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
H2S – Hydrogen sulfide |
|||
Electrochemical sensor |
1 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
H2 – Hydrogen |
|||
Electrochemical sensor |
2 000ppm | 1ppm |
± 10ppm abs. or 5% rel |
45 sec |
Electrochemical sensor |
20 000ppm | 1ppm |
± 10ppm abs. or 5% rel |
45 sec |
NH3 – Ammonia |
|||
Electrochemical sensor |
1 000ppm | 1ppm |
± 10ppm abs. or 5% rel |
45 sec |
CL2 - Chlorine |
|||
Electrochemical sensor |
250ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
HCl - Hydrogen chloride |
|||
Electrochemical sensor |
50ppm | 1ppm |
± 5ppm abs. or 5% rel |