Sensonic IR-1 Gas Analyzers |
|||
Specifications |
|||
Dimensions ( W x H x D ) |
500 mm x 395 mm x 173 mm |
||
Weight |
14 kg ÷ 18 kg |
||
Casing's material |
Plywood covered with aluminium |
||
Operating conditions |
T: 10°C ÷ 50°C, RH: 5% ÷ 90% ( non-condensing ) |
||
Storing temperature |
-20°C ÷ +55°C |
||
Power supply : input | maximal power consumption |
115 or 230VAC | 150W |
||
Operating system |
Windows CE 5.0 |
||
Display |
6.4” VGA ( 640 x 480 ) |
||
Data storage: type | capacity |
Compact Flash card | max. 4GB |
||
Interface for external devices ( USB disk, mouse, keyboard ) |
2x USB |
||
Communication interface with PC computer |
RS-232C, RJ45 ( Ethernet ) |
||
Warming up time |
90min max |
||
Warming up temperature |
about 18°C above ambient temperature |
||
Maximum outside temp. drift that does not affect warming up temp. |
±5°C |
||
Measurements of gas components |
|||
Method |
Range | Resolution |
Accuray |
Time (T90) |
O2 – Oxygen |
|||
Electrochemical sensor |
20.95% | 0.01% |
± 0.1% abs. or 5% rel |
45 sec |
Electrochemical, partial pressure |
20.95% | 0.01% |
± 0.1% abs. or 5% rel |
45 sec |
Electrochemical, partial pressure |
25% | 0.01% |
± 0.1% abs. or 5% rel |
45 sec |
Electrochemical, partial pressure |
100% | 0.1% |
± 0.1% abs. or 5% rel |
45 sec |
Paramagnetic sensor |
25% | 0.01% |
± 0.1% abs. or 5% rel |
45 sec |
Paramagnetic sensor |
100% | 0.1% |
± 0.1% abs. or 5% rel |
45 sec |
CO – Carbon monoxide |
|||
Electrochemical sensor |
20 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
Electrochemical, with H2 compensation |
2 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
Electrochemical sensor |
10% | 10ppm |
± 50ppm abs. or 5% rel |
45 sec |
NDIR |
20 000ppm | 1ppm |
± 3ppm abs. or 3% rel |
45 sec |
NDIR |
10% | 0.01% |
± 0.3% abs. or 3% rel |
45 sec |
NDIR |
100% | 0.1% |
± 0.3% abs. or 3% rel |
45 sec |
CO2 – Carbon dioxide |
|||
NDIR |
5% | 0.01% |
± 0.3% abs. or 3% rel |
45 sec |
NDIR |
10% | 0.01% |
± 0.3% abs. or 3% rel |
45 sec |
NDIR |
25% | 0.01% |
± 0.3% abs. or 3% rel |
45 sec |
NDIR |
50% | 0.01% |
± 0.3% abs. or 3% rel |
45 sec |
NDIR |
100% | 0.1% |
± 0.3% abs. or 3% rel |
45 sec |
CxHy – Total Hydrocarbons |
|||
NDIR |
5% | 0.1% |
± 0.3% abs. or 3% rel |
45 sec |
NDIR |
10% | 0.1% |
± 0.3% abs. or 3% rel |
45 sec |
NDIR |
25% | 0.1% |
± 0.3% abs. or 3% rel |
45 sec |
NDIR |
50% | 0.1% |
± 0.3% abs. or 3% rel |
45 sec |
NDIR |
100% | 0.1% |
± 0.3% abs. or 3% rel |
45 sec |
NO – Nitric oxide |
|||
NDIR |
5 000ppm | 1ppm |
± 3ppm abs. or 3% rel |
45 sec |
Electrochemical sensor |
5 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
NO2 – Nitrogen dioxide |
|||
NDIR |
1 000ppm | 1ppm |
± 3ppm abs. or 3% rel |
45 sec |
Electrochemical sensor |
5 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
SO2 – Sulphur dioxide |
|||
NDIR |
5 000ppm | 1ppm |
± 3ppm abs. or 3% rel |
45 sec |
Electrochemical sensor |
5 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
H2S – Hydrogen sulfide |
|||
Electrochemical sensor |
1 000ppm | 1ppm |
± 5ppm abs. or 5% rel |
45 sec |
H2 – Hydrogen |
|||
Electrochemical sensor |
2 000ppm | 1ppm |
± 10ppm abs. or 5% rel |
45 sec |
Electrochemical sensor |
20 000ppm | 1ppm |
± 10ppm abs. or 5% rel |
45 sec |
NH3 – Ammonia(1) |
|||
Electrochemical sensor |
1 000ppm | 1ppm |
± 10ppm abs. or 5% rel |
45 sec |
N2O – Nitrous oxide |
|||
NDIR |
2 000ppm | 1ppm |
± 3ppm abs. or 3% rel |
45 sec |
CHF3 – Fluoroform (Refrigerant R23) |
|||
NDIR |
2.5% | 0.01% |
± 0.3% abs. or 3% rel |
45 sec |
Notes : (1) – not available with dryer equipped only with Nafion® exchanger as NH3 is destructive for nafion |
|||
Other measurements / calculated results |
|||
Method |
Range | Resolution |
Accuracy |
Time (T90) |
Tgas – gas temperature |
|||
K-type thermocouple |
-10 ÷ 1000°C | 0.1°C |
± 2°C |
10 sec |
S-type thermocouple |
-10 ÷ 1500°C | 0.1°C |
± 2°C |
10 sec |
Tamb – boiler intake air temperature |
|||
PT500 resistive sensor |
-10 ÷ 100°C | 0.1°C |
± 2°C |
10 sec |
Differential pressure |
|||
Silicon piezoresistive pressure sensor |
-25hPa ÷ +25hPa | 1Pa (0.01hPa) |
± 2Pa abs. or 5% rel |
10 sec |
Gas flow velocity |
|||
Indirect, with Pitot tube & pressure sensor |
1 ÷ 50m/s | 0.1m/s |
0.3m/s abs. or 5% rel |
10 sec |
Lambda λ – excess air number |
|||
Calculated |
1 ÷ 10 | 0.01 |
± 5% rel |
10 sec |
qA – stack loss |
|||
Calculated |
0 ÷ 100% | 0.1% |
% ± 5% rel |
10 sec |
Eta η – combustion efficiency |
|||
Calculated |
0 ÷ 120% | 0.1% |
± 5% rel |
10 sec |
IL – incomplete combustion |
|||
Calculated |
0 ÷ 100% | 0.01% |
± 5% rel |