Parker icountBSplus - Bottle Sampler Kittiwake

Specifications

Principle of operation

Laser diode optical detection of actual particulates

Dimensions

H 530 mm x W 210 mm x D 410 mm

Weight

Approx 18 kg

Operating temp. & humidity

+5°C to +60°C ( -41°F to +140°F )

20 - 85% RH ( tested at 30°C (86°F), non-condensing )

Storage temp. & humidity

-40°C to +90°C ( -40°F to +194°F )

10 - 90% RH ( tested at 30°C (86°F), non-condensing )

Moisture sensor calibration

± 5% RH (over a compensated temperature range of +10°C to +80°C (+50°F to +176°F)

Moisture sensor stability 

± 2% RH typical at 50% RH in one year

International codes

ISO 7 to 21, NAS 0 to 12, AS 0 to 12

Contamination standards

Refer to Parker 'Guide to Contamination Standards' (DD0000015) on CD

MTD: ISO 4406:1999; NAS 1638; AS4059E (Differential); AS4509E (Cumulative)

ACFTD: ISO 4406:1987; ISO4406:1991; NAS 1638; AS4509E (Differential); AS4509E (Cumulative); GOST 17216:2001

Calibration

 MTD: via a certified primary ISO 11171 automatic particle detector using ISO 11943 principles, with particle distribution reporting to ISO 4406:1996

ACFTD: fully traceable to gravimetric first principles

Recalibration

Contact Parker Hannifin for advice

Fluid compatibility

Mineral-based oils and petroleum-based fuel. Contact Parker Hannifin for advice

Fluid management

Selectable on screen between 25 m to 100 ml

Viscosity range

1 to 300 cSt

Working pressure

3 bar maximum input pressure, if used on-line. Contact Parker Hannifin for further advice

Flow range through icountBSplus

Test: 60 ml/min

Connection interface (on-line)

INLET: 6 mm push-fit, DRAIN: 4 mm push-fit

Fluid operating temp. (oil)

+5°C to +70°C (-41°F to 176°F)

Fluid operating temp.(fuel)

+20°C to +70°C (-4°F to 158°F)

Sample bottle size

See Parker ACC Spares list. Contact Parker Hannifin for advice

Flush sample size

Selectable option within the icountBS software: 25 ml to 100 ml

Memory storage

500 tests (Integrated Warning Level)

Printer

Thermal dot line printer - see ACC spares list for replacement paper

Battery type

Polymer Lithium Ion Battery pack (ACC6NW032)

Power requirements

Integrated supply into the icountBSplus unit

Certification

CE Certified. Supplied with EC Declaration of Conformity Certificate

Ordering Information

IBS3000

iBSplus (online)

IBS3100

iBSplus (offline)