Parker icount Oil Sampler ( IOS ) Kittiwake |
|
Specifications |
|
Product start-up time |
10 seconds minimum |
Measurement period |
Default 30 seconds run time; 15 seconds data logging time |
Reporting interval |
On-board data storage every second. Output via RJ45 connection |
Principle of operation |
Laser diode optical detection of actual particulates |
International codes range |
Up to ISO 22 ( +/- 1 ISO code ) NAS 0 - 12 |
Calibration |
Calibration by recognised on-line methods confirmed by the relevant ISO procedures. MTD - via a certified primary ISO 11171 automatic particle detector using ISO 11943 principles. Particle distribution reporting to ISO 4406:1999 |
Recalibration and Servicing |
Recommended every 12 months |
Working pressure |
2.5 - 350 bar ( 35 - 5000 psi ). Pressures above 2.5 bar require the use of a Parker Pressure Reducing Valve (PRV) - ACC6NN027 |
Working viscosity |
1 - 300 cSt |
Flow range through IOS |
40 - 140 ml/minute; controlled at 60 ml/min by IOS's internal pump |
Fluid connection interface |
INLET: 6 mm push-fit. DRAIN: 4mm push-fit |
Ambient storage temp. for unit |
-40°C to +80°C; -40°F to +176°F |
Operating temperature for unit |
-30°C to +80°C; -22°F to +176°F |
Operating humidity range |
5% RH to 100% RH |
Fluid operating temp. (oil) |
+5°C to +80°C; +41°F to +176°F |
Fluid operating temp. (fuel) |
-20°C to +70°C; -4°F to +158°F |
Moisture sensor |
Linear scale within the range 5% RH to 100% RH |
Computer compatibility |
IP68-rated RJ45 connection that may be connected to a laptop computer's RJ45 LAN port using the 2 m cable supplied |
Power requirement |
Regulated power supply supplied with the unit |
Certification |
IP54 rating ( unit open ) |
IP67 rating ( unit closed ) |
|
EC Declaration of Conformity |
|
Machinery Directive |
|
EMC EN61000-6-3:2001 |
|
EMC EN61000-6-2:2001 |
|
EMC EN61010-1:2001 |
|
CE Certified |
|
Ordering Information |
|
IOS1220UK |
IOS, Mineral MTD online UK |
IOS1210UK |
IOS, Mineral MTD Offline UK |
IOS1221UK |
IOS Mineral Wireless online UK |