Humidity Chamber HCP108, HCP153 & HCP246 Memmert |
|
Control of standard components |
|
Controller |
Electronic microprocessor temperature controller with auto-diagnostic system |
Temperature |
2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value |
Timer |
Digital 7-day programme timer with real time clock, precise minute setting |
Timer |
Integrated timer for tempering profiles of up to 40 ramps each, each segment adjustable from 1 min. To 999 hrs. |
Controller |
Digital display of all set parameters, such a temperature, weekdays, time, CO2, humidity and set-up values - language to be chosen via set-up |
Humidity |
Active humidifying and de-humidifying control ( 20-95 % ) with digital display of relative humidity - resolution of display 0.5 %, setting accuracy 1 % |
Humidity |
Humidity supply with distilled water from external tank by self-priming pump |
Humidity |
Active humidifying and de-humidifying adjustable from 20-95 % rh with digital display of relative humidity - resolution of display 0.5 %, setting accuracy 1 % |
Humidity |
Humidity supply with distilled water from external tank by self-priming pump |
Humidity |
Humidification by hot steam generator |
Temperature |
Resolution of display for setpoint values 0.1°C up to 99.9°C, 0.5°C from 100°C and for actual values |
With humidity min. 8°C above ambient up to +90°C |
|
Without humidity min. 8°C above ambient up to +160°C |
|
Control technology |
|
Calibration |
Three freely selectable temperature values |
Ventilation |
Uniform atmosphere and temperature distribution owing to enclosed non-turbulent ventilation system in working chamber |
Communication |
|
documentation |
Integrated ring memory as data logger for GLP-conforming long-term documentation of all relevant parameters - 1024 kB |
documentation |
Programme stored in case of power failure |
documentation |
Parallel printer interface ( incl. real time clock with date function ) for all PCL3-compatible ink jet printers for GLP-conforming documentation |
Interface |
USB-interface incl. Memmert software "Celsius" for programming and documentation |
Programming |
Chip-card control incl. 1 MEMoryCard XL with 32 kB storage capacity (max. 40 ramps) |
Safety |
|
Autodiagnostic system |
Integral fault diagnostics for temperature and humidity control |
Alarm |
Visual and acoustic |
Temperature control |
Overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display |
Alarm |
With visual and acoustic alarm in case of over-/under temperature and underhumidity, open door and empty water tank |
AutoSAFETY |
Additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature |
Temperature control |
Mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 10 °C above nominal temperature |
Heating concept |
|
Large-area multi-function heating system on four sides with additional door and back heating to avoid condensation |
|
Standard equipment |
|
Scope of delivery |
2nd chip-card ( STERICard ) for sterilisation of working chamber with fixed values ( 4 hours/160°C ) without removal of sensors |
Scope of delivery |
Incl. works calibration certificate for +60°C |
Door |
Fully insulated stainless steel door with2-point locking (compression door lock), lockable |
Door |
Inner glass door |
Housing |
Rear zinc-plated steel |
Interior |
Easy-to-clean interior,made of stainless steel,reinforced by deep drawn ribbing, material 1.4301 (ASTM 304), hermetically welded |
Internals |
2 perforated stainless steel shelves |
HUMIDY CHAMBER HCP 108 |
|
Stainless steel interior |
|
Dimensions W x H x D |
w(A) x h(B) x d(C) : 560 x 480 x 400 mm |
Volume |
108 Liter |
Textured stainless steel casing |
w(D) x h(E) x d(F) : 710 x 778 x 550 mm |
Electrical data |
|
Voltage |
230 V, 50/60 Hz |
Electrical load |
Approx. 1000 W |
Dimensions approx incl. carton |
B x H x T: 830 x 1050 x 800 mm |
Net weight |
Approx. 70 kg |
Gross weight carton |
Approx. 95 kg |
HUMIDY CHAMBER HCP 153 |
|
Stainless steel interior |
|
Dimensions W x H x D |
w(A) x h(B) x d(C) : 480 x 640 x 500 mm |
Volume |
153 Liter |
Textured stainless steel casing |
w(D) x h(E) x d(F) : 630 x 938 x 650 mm |
Electrical data |
|
Voltage |
230 V, 50/60 Hz |
Electrical load |
Approx. 1500 W |
Dimensions approx incl. carton |
B x H x T : 830 x 1300 x 800 mm |
Net weight |
Approx. 80 kg |
Gross weight carton |
Approx. 106 kg |
HUMIDY CHAMBER HCP 246 |
|
Stainless steel interior |
|
Dimensions W x H x D |
w(A) x h(B) x d(C) : 640 x 640 x 600 mm |
Volume |
246 Liter |
Textured stainless steel casing |
w(D) x h(E) x d(F) : 790 x 938 x 750 mm |
Electrical data |
|
Voltage |
230 V, 50/60 Hz |
Electrical load |
Approx. 2000 W |
Dimensions approx incl. carton |
B x H x T: 930 x 1140 x 930 mm |
Net weight |
Approx. 110 kg |
Gross weight carton |
Approx. 132 kg |