Gas Leak Checker Detector SP-220 Riken Keiki |
||||
Specification for Type FUM & SC |
||||
Model |
SP-220 TYPE FUM |
SP-220 TYPE SC |
||
( Fumigation ) |
( Toxic, Combustible, CFC ) |
|||
Target gas |
Refer to the gas list |
|||
Detection principle |
Hot-wire semiconductor method |
|||
Sensor |
SH-8662 |
|||
Measuring range |
Depends on target gas |
|||
Alarm set point |
Depends on target gas |
|||
Accuracy |
Displays 0.1 ppm-bar when applied PH3 0.3 ppm. |
|||
Alarm pattern |
Gas alarm : Gas concentration reached or over the setting point. |
|||
Trouble alarm : Sensor connection / breaking, Low battery, Low flow Circuit error, Calibration error |
||||
Alarm method |
Non-latching |
|||
Display of alarm |
Gas alarm : Flashing LED, Buzzer |
|||
Trouble alarm : Flashing LED, Buzzer, Display of error message |
||||
Alarm delay time |
0 sec. |
|||
Sampling method |
Sample draw |
|||
Response time |
10 sec.: Time to reach 0.1 ppm when applied PH3 0.3 ppm. |
|||
Output signal |
IrDA |
|||
Switch |
「AIR」「MODE」「POWER」 |
|||
Display device |
LCD |
|||
Display contents |
Gas name, Gas concentration, Time, Battery, Pilot indicator, Pump driving indicator |
|||
Power source |
2 AA dry cell batteries |
|||
Continuous operation |
More than 12 hrs. |
|||
Operating temp. & humidity |
-20~+55 °C ( -4 °F±131 °F ), 0~95 % RH( Non-condensing C30 ) |
|||
Dimension & weight |
200 H x 43 D x 47 mm W ( Taper nozzle not included ), 215 g |
|||
Ingress proof |
IP 55 |
|||
Function |
Manual LCD backlight ( Automatically ON when alarming ), Peak value, Snap logging, Time, Cal alarm |
|||
Specification for Type M, L & ML |
||||
Model |
SP-220 ( TYPE M ) |
SP-220 ( TYPE L ) |
SP-220 ( TYPE ML ) |
|
Detection principle |
Hot-wire semiconductor method |
|||
Target gas |
Town gas ( switchable to LPG ) |
LPG ( switchable to town gas ) |
Town-gas, LPG ( switching ) |
|
|
*2 gases are needed for calibration |
|||
Concentration display |
LCD level meter + Scale |
|||
Measuring range |
10~10000 ppm |
|||
Sampling method |
Sample draw |
|||
Alarm set point |
Default:30 ppm ( selectable from 10, 30, 150, 500, 2000 ppm ) |
|||
Display |
Operating state disp., Flow check disp., Gas concentration disp., Alarm sound disp., Operation sound disp., Battery level icon, Level meter disp., Mode disp., Gas name and message disp. |
|||
Response time |
3 sec. |
|||
Gas alarm display |
Flashing LED, Buzzer |
|||
Gas alarm method |
Non-latching |
|||
Trouble alarm |
Sensor error, Low battery, Low flow, System error, Calibration error, Clock error, Pump error |
|||
Trouble alarm display |
Flashing LED, Buzzer, Trouble message |
|||
Trouble alarm method |
Latching |
|||
Power source |
2 AA dry cell batteries |
|||
Continuous operation |
Approx. 13 hrs. |
|||
Operating temperature |
-20~55 °C ( -4 °F~131 °F C48 ) |
|||
Operating humidity |
0~95 % RH ( Non-condensing ) |
|||
Safety design |
Intrinsically safety ExiaIICT4 |
|||
Approvals |
Ex-proof : IECEx, ATEX / Other: CE marking |
|||
Ingress proof |
IP55 |
|||
Dimension |
200 ( H ) × 43 ( W ) × 39 ( D ) mm ( Taper nozzle not included ) |
|||
Weight |
215 g ( Dry cell batteries not included ) |