Autoclave ALP KT-40 & KT-40DP |
|||||||
Specifications |
|||||||
Model |
Sterilization only |
KT-23 |
KT-30S |
KT-30L |
KT-40S |
KT-40 |
KT-40L |
With warm air drying |
KT-23DP |
KT-30SDP |
KT-30LDP |
KT-40SDP |
KT-40DP |
KT-40LDP |
|
Chamber ( inner dia x depth ) |
23 ¢ X 46 cm |
30 ¢ X 45 cm |
30 ¢ X 64 cm |
40 ¢ X 50 cm |
40 ¢ X 65 cm |
40 ¢ X 83 cm |
|
Usable capacity |
20 L |
32 L |
45 L |
62 L |
82 L |
105 L |
|
Ability |
Usable max. press. |
0. 1.6 MPa ( 1.6 kg/cm² ) |
|||||
Usable max. temp. |
127 ºC |
||||||
Sterilization |
110-127 ºC |
||||||
One-touch setting temp. |
110, 115, 121, 127 ºC |
||||||
Sterilization System |
System |
Saturated steam generated by sheathed electric heater |
|||||
Main controller |
Precise pressure switch |
||||||
Temp. set system |
Analog temp. set on the panel of pressure switch |
||||||
Temp. indicator |
Saturated steam scale on the panel of pressure gauge |
||||||
Timer |
1 ~ 60 min. ( Operates after the set temp. is reached.) |
||||||
Operation start |
PUSH-ON start button switch |
||||||
Air exhaust valve |
Automatic exhaust system of remained air by heat expansion bellows valve |
||||||
Water lack preventor |
Automatic cut-off system by a sensitive thermostat ( Warning by buzzer and lamp ) |
||||||
|
Exhaust valve |
Manual needle valve being opened only in required rapid exhaust Drain valve Manual ball valve ( Inner dia. : 10 mm ) |
|||||
Sterilization pattern |
PUSH-ON start → Heating, Air exhaust → Sterilization → Completion buzzer → Exhaust or natural cooling |
||||||
KT-DP Models |
Operation pattern with drying |
PUSH-ON start → Heating, Air exhaust → Sterilization → Completion buzzer |
|||||
Open DRAIN valve ( by manual ) while pressure remains → Pressure goes down to zero |
|||||||
Switching from STER. to DRY ( by manual ) → Set drying time → PUSH-ON drying start → Drying |
|||||||
→ Drying completion buzzer |
|||||||
Drying devices |
Warm air circulating and discharge by electric heater and air pump through 0.2 µm HEPA cartridge filter |
||||||
Drying temp. : 130 ºC ( changeable by inner dia l ), Timer 0 - 60 min |
|||||||
Safety |
Safety devices |
Water lack preventer. Saftey valve |
Water lack preventer, Saftey valve |
||||
Over pressure saftey switch |
|||||||
Electric safety |
Breaker for over current |
Breaker for over current & slight leakage |
|||||
Withstand valtage : 1.500 V, 1 min. Insulation : 100 MΩ- ( 500 V.D.C ) |
|||||||
Safety valve operation |
0.18 MPa |
||||||
Water test pressure |
0.36 MPa |
||||||
Material |
Chamber & lid |
Stainless steel SUS304 shiny polished. Lid gasket : Molded silicon rubber |
|||||
Outer panel |
Upper surface : SUS443J 1, Main body : Steel with baked enamel,Caster ( front two with stopper ) |
||||||
Power requirement |
AC 110 / 120 / 220 / 230 / 240 V, 1Ø 50/60 Hz |
||||||
( 50/60 Hz ) |
1.5 kW |
2.0 kW |
4.0 kW |
||||
Outer size W x D x H ( cm ) |
53 X 40 X 87 |
60 X 40 X 87 |
60 X 40 X 95 |
73 X 57 X 76 |
73 X 57 X 91 |
73 X 57 X 109 |
|
Weight |
35/40 kg |
49/54 kg |
53/58 kg |
67/76 kg |
72/81 kg |
86/95 kg |