Description | |
Type of chamber, oven:
|
Incubator |
Incubator model: | CO2 Incubator |
Inner volume: | 101 - 200 l |
Maximum load: | 11 - 50 kg |
Highest temperature ( heat ): | +60 °C |
Width inside ( incubators,...): | 500 mm |
Depth inside ( incubators,...): | 500 mm |
Height inside ( incubators,...): | 600 mm |
Width outside ( incubators,...): | 680 mm |
Depth outside ( incubators,...): | 715 mm |
Height outside ( incubators,...): | 920 mm |
Technical Data | |
BINDER CO2 incubator | CB 160 |
Exterior dimensions | |
Width ( mm ) | 680 |
Height ( incl. Feet ) ( mm ) | 920 |
Depth ( plus 54 mm for door handle and connection ) ( mm ) | 715 |
Wall clearance side / rear ( mm ) | 50 / 100 |
Number of doors ( ea.) | 1 |
Inner glass door ( s ) | 1 |
Interior dimensions | |
Width ( mm ) | 500 |
Height ( mm ) | 600 |
Depth ( mm ) | 500 |
Interior volume ( l ) | 150 |
Stainless steel shelf ( number standard/max.) | 3/6 |
Dimensions of shelf, width x depth ( mm ) | 495 x 444 |
Load per shelf ( kg ) | 10 |
Permitted total load ( kg ) | 30 |
Weight ( empty ) ( kg ) | 107 |
Temperature data | |
Temperature range from 7 °C above ambient temperature to ( °C ) | 60 |
Temperature variation at 37 °C ( ± K ) | 0,3 |
Temperature fluctuation at 37 °C ( ± K ) | 0,1 |
Recovery time after door was opened for 30 sec. at 37 °C ( min.) 1 | 4 |
Humidity data | |
Humidity ( ±2 % r.h.) | 90-95 |
CO2 data | |
CO2 range ( Vol.-% CO2 ) | 0-20 |
Setting accuracy ( Vol.-% CO2 ) | 0,1 |
Recovery time after door was opened for 30 sec. 1) | 5 |
at 5.0 vol. % CO2 ( min.) | |
CO2 measurement | IR |
Connection hose nozzle for CO2 DN 6 for hose with internal diameter ( mm ) | 6 |
O2 data | |
O2 range ( Vol.-% O2 ) | 0,2 - 95 |
Setting accuracy ( Vol.-% O2 ) | 0,1 |
Recovery time after door was opened for 30 sec. 1) | |
at 1.0 vol. % O2 ( min.) | 12 |
at 5.0 vol. % O2 ( min.) | 12 |
O2 measurement | ZrO2 |
Connection hose nozzle for O2 / N2 DN 6 for hose with internal diameter ( mm ) | 6 |
Electrical data | |
Voltage ( ± 10 % ) | |
50/60 Hz ( V ) | 230 |
60 Hz ( V ) | 115 |
Nominal power ( kW ) | 1,3 |
Energy consumption at 37 °C ( W ) 2) | 100 |
1) up to 98 % of the set value | |
2) These values can be used for dimensioning air condition systems. |