CO2 Incubator CB 160 Binder
Description
Type of chamber, oven:
Incubator
Incubator model: CO2 Incubator
Inner volume: 101 - 200 l
Maximum load: 11 - 50 kg
Highest temperature ( heat ): +60 °C
Width inside ( incubators,...): 500 mm
Depth inside ( incubators,...): 500 mm
Height inside ( incubators,...): 600 mm
Width outside ( incubators,...): 680 mm
Depth outside ( incubators,...): 715 mm
Height outside ( incubators,...): 920 mm
Technical Data
BINDER CO2 incubator CB 160
Exterior dimensions
Width ( mm ) 680
Height ( incl. Feet ) ( mm ) 920
Depth ( plus 54 mm for door handle and connection ) ( mm ) 715
Wall clearance side / rear ( mm ) 50 / 100
Number of doors ( ea.) 1
Inner glass door ( s ) 1
Interior dimensions
Width ( mm ) 500
Height ( mm ) 600
Depth ( mm ) 500
Interior volume ( l ) 150
Stainless steel shelf ( number standard/max.) 3/6
Dimensions of shelf, width x depth ( mm ) 495 x 444
Load per shelf ( kg ) 10
Permitted total load ( kg ) 30
Weight ( empty ) ( kg ) 107
Temperature data
Temperature range from 7 °C above ambient temperature to ( °C ) 60
Temperature variation at 37 °C ( ± K ) 0,3
Temperature fluctuation at 37 °C ( ± K ) 0,1
Recovery time after door was opened for 30 sec. at 37 °C ( min.) 1 4
Humidity data
Humidity ( ±2 % r.h.) 90-95
CO2 data  
CO2 range ( Vol.-% CO2 ) 0-20
Setting accuracy ( Vol.-% CO2 ) 0,1
Recovery time after door was opened for 30 sec. 1) 5
at 5.0 vol. % CO2 ( min.)  
CO2 measurement IR
Connection hose nozzle for CO2 DN 6 for hose with internal diameter ( mm ) 6
O2 data
O2 range ( Vol.-% O2 ) 0,2 - 95
Setting accuracy ( Vol.-% O2 ) 0,1
Recovery time after door was opened for 30 sec. 1)
at 1.0 vol. % O2 ( min.) 12
at 5.0 vol. % O2 ( min.) 12
O2 measurement ZrO2
Connection hose nozzle for O2 / N2 DN 6 for hose with internal diameter ( mm ) 6
Electrical data
Voltage ( ± 10 % )
50/60 Hz ( V ) 230
60 Hz ( V ) 115
Nominal power ( kW ) 1,3
Energy consumption at 37 °C ( W ) 2) 100
1) up to 98 % of the set value 
2) These values can be used for dimensioning air condition systems. 